DEFIBRILLATOR DF RELIFE 900/R
The Relife 900 is a biphasic defibrillator system that provides advanced monitoring functions and acute cardiac care by multiple modes of operation: AED, manual defibrillation, synchronized cardioversion, external pacing and monitoring - MASIMO SPO2, SunTech NIBP, ETCO2.
Features
7 inch Colour TFT Screen
Allows the clear and precise display of multiple waveforms and defibrillation parameters
Bi-phasic waveform (Current - Controlled)
The current controlled bi-phasic waveforms lowers the electrical threshold for successful defibrillation, besides adjusting for impedance
High Energy Efficiency
100 charge / discharge of 300J with fully charged new battery
Wide Energy Levels
2-300 J of energy level can be selected for defibrillation by the user
Remote energy selection and charging
Remote energy selection, charging and discharging through paddles facilitate convenient usage
Voice Prompt Support
Continuous voice prompts and on screen messages to aid in revival and resuscitation process
Metronome Signal for CPR Support
Metronome signal for synchronizing chest compression during CPR
Synchronized Cardioversion
Synchronized Cardioversion function is available for providing defibrillation on sync with ECG waveform
24 Events
Memorises always the last 24 Critical Event ECG information which can be viewed or recorded
Patient-Paddle Contact Indication
Patient-Paddle quality of contact is provided on screen
Technical Specifications |
|
Manual Defibrillator |
|
Defibrillator Waveform |
Current Controlled, Biphasic |
Energy Selection Option |
2-300 Joules in steps of 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 250, 300 Joules |
Energy Selection Control |
Selected using keys located on the front panel or on the sternum paddle |
Charge Control |
Control on the front panel and on the apex paddle |
Charge Ready Indication |
Charge done LED on the front panel CHARGE key and Charge done LED on the apex paddle |
Synchroniser |
Energy is delivered within 50 msec of the detected R-wave peak |
Patient Impedance |
25 – 200 Ohms |
AED |
|
AED Mode Selection Control |
Control on the front panel |
Shock Sequence |
Low-Low-Low / Low-Low-High / Low-High-High (User selectable) |
Charge Ready Indication |
Onscreen display, voice prompt and blinking LED in the SHOCK key |
Shock Delivery Control |
Control on the front panel with blinking LED |
ECG Analysis Time |
< 15 seconds |
ECG |
|
ECG Input Modes |
ECG input through Paddles or ECG Patient cable |
ECG Lead Selection |
LEAD I / II / III / PADDLES (3 Lead Patient cable) |
LEAD I / II / III / aVR / aVL / aVF / V / PADDLES (5 Lead Patient cable |
|
CMRR |
>90 dB at 50 Hz |
Frequency Response |
0.5 to 40 Hz |
Heart Rate Display |
10 to 300 bpm; Accuracy ± 2 bpm |
User Gain Settings |
8 level setting 2.5, 5, 7.5, 10, 15, 20, 30 and 40 mm/mV |
Sweep Speed |
25 ± 1 mm/sec |
Reset Recovery |
Automatic recovery of waveform within 0.5 seconds after overload or defibrillation |
MASIMO SPO2 |
|
Spo2 Range |
1 – 100% |
Accuracy |
±2% (adult/child, in non-motion state), ±3% (adult/child, in motion state) within the measurement range of 70%~100%. 1%~69% is not defined. |
Averaging Time |
2-4 / 4-6 / 8 / 10 / 12 / 14 / 16 Seconds (User selectable) |
Resolution |
1% |
Pulse Rate Rang |
25 – 240 bpm |
Perfusion Index Resolution |
0.01% |
SunTech NIBP |
|
Method |
Oscillometric |
Operating Modes |
Manual, Automatic and Stat |
Blood Pressure Range |
|
Systolic |
Adult: 40 – 250 mmHg | Paediatric: 40 – 230 mmHg |
Diastolic |
Adult: 20 – 200 mmHg | Paediatric: 20 – 160 mmHg |
MAP |
Adult: 26 – 220 mmHg | Paediatric: 26 – 183 mmHg |
Pressure Transducer Accuracy |
± 3 mmHg or 2% of the reading whichever is greater |
Cycle Time in Auto mode |
1, 2, 3, 5, 10, 15, 30, 60, 90 Minutes (User selectable |
Cycle Time in Stat mode |
5 Minutes |
Thermal Recorder |
|
Paper Speed |
25 ± 1 mm/sec |
Paper Size And Type |
Thermal paper in rolls. 58 mm width, 15 meters |
Print Channel |
1 |
General |
|
Display |
7” Colour TFT, Resolution 800 x 480 |
Dimensions |
370 mm (W) x 300 mm (H) x 290 mm (D) |
Weight |
Less than 8.5 Kg. (including battery, external paddles) |
Operating Temperature Range |
0 to 40 ˚C at non-condensing RH between 15 and 95% |
Storage Temperature Range |
-10 to 60˚C at non-condensing RH between 20 and 80% |
AC Input |
100 – 240 Vac, 50/60 Hz |
Power Consumption |
< 240 VA |
Battery Type |
Sealed Lead Acid 12V, 4.5 AH, Rechargeable |
Battery Capacity |
> 100 discharges of 300 Joules |
Battery Charging Time |
Approx. 3 hours to 90% of full capacity. 18 hours to 100% capacity |
Event Information |
Last 24 critical events are stored along with ECG information. Events include MARK key press, Shock delivery and HR Alarm violation. |
Trend |
Last 200 NIBP measurement data and last 200 Alarm violation information |
SKU-MXPT_SV-SNM
Estimated Delivery Time:_5 to 7 Working Days
VARIANT | SELLER | PRICE | QUANTITY |
---|
Brand | BPL |
DEFIBRILLATOR DF RELIFE 900/R
The Relife 900 is a biphasic defibrillator system that provides advanced monitoring functions and acute cardiac care by multiple modes of operation: AED, manual defibrillation, synchronized cardioversion, external pacing and monitoring - MASIMO SPO2, SunTech NIBP, ETCO2.
Features
7 inch Colour TFT Screen
Allows the clear and precise display of multiple waveforms and defibrillation parameters
Bi-phasic waveform (Current - Controlled)
The current controlled bi-phasic waveforms lowers the electrical threshold for successful defibrillation, besides adjusting for impedance
High Energy Efficiency
100 charge / discharge of 300J with fully charged new battery
Wide Energy Levels
2-300 J of energy level can be selected for defibrillation by the user
Remote energy selection and charging
Remote energy selection, charging and discharging through paddles facilitate convenient usage
Voice Prompt Support
Continuous voice prompts and on screen messages to aid in revival and resuscitation process
Metronome Signal for CPR Support
Metronome signal for synchronizing chest compression during CPR
Synchronized Cardioversion
Synchronized Cardioversion function is available for providing defibrillation on sync with ECG waveform
24 Events
Memorises always the last 24 Critical Event ECG information which can be viewed or recorded
Patient-Paddle Contact Indication
Patient-Paddle quality of contact is provided on screen
Technical Specifications |
|
Manual Defibrillator |
|
Defibrillator Waveform |
Current Controlled, Biphasic |
Energy Selection Option |
2-300 Joules in steps of 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 250, 300 Joules |
Energy Selection Control |
Selected using keys located on the front panel or on the sternum paddle |
Charge Control |
Control on the front panel and on the apex paddle |
Charge Ready Indication |
Charge done LED on the front panel CHARGE key and Charge done LED on the apex paddle |
Synchroniser |
Energy is delivered within 50 msec of the detected R-wave peak |
Patient Impedance |
25 – 200 Ohms |
AED |
|
AED Mode Selection Control |
Control on the front panel |
Shock Sequence |
Low-Low-Low / Low-Low-High / Low-High-High (User selectable) |
Charge Ready Indication |
Onscreen display, voice prompt and blinking LED in the SHOCK key |
Shock Delivery Control |
Control on the front panel with blinking LED |
ECG Analysis Time |
< 15 seconds |
ECG |
|
ECG Input Modes |
ECG input through Paddles or ECG Patient cable |
ECG Lead Selection |
LEAD I / II / III / PADDLES (3 Lead Patient cable) |
LEAD I / II / III / aVR / aVL / aVF / V / PADDLES (5 Lead Patient cable |
|
CMRR |
>90 dB at 50 Hz |
Frequency Response |
0.5 to 40 Hz |
Heart Rate Display |
10 to 300 bpm; Accuracy ± 2 bpm |
User Gain Settings |
8 level setting 2.5, 5, 7.5, 10, 15, 20, 30 and 40 mm/mV |
Sweep Speed |
25 ± 1 mm/sec |
Reset Recovery |
Automatic recovery of waveform within 0.5 seconds after overload or defibrillation |
MASIMO SPO2 |
|
Spo2 Range |
1 – 100% |
Accuracy |
±2% (adult/child, in non-motion state), ±3% (adult/child, in motion state) within the measurement range of 70%~100%. 1%~69% is not defined. |
Averaging Time |
2-4 / 4-6 / 8 / 10 / 12 / 14 / 16 Seconds (User selectable) |
Resolution |
1% |
Pulse Rate Rang |
25 – 240 bpm |
Perfusion Index Resolution |
0.01% |
SunTech NIBP |
|
Method |
Oscillometric |
Operating Modes |
Manual, Automatic and Stat |
Blood Pressure Range |
|
Systolic |
Adult: 40 – 250 mmHg | Paediatric: 40 – 230 mmHg |
Diastolic |
Adult: 20 – 200 mmHg | Paediatric: 20 – 160 mmHg |
MAP |
Adult: 26 – 220 mmHg | Paediatric: 26 – 183 mmHg |
Pressure Transducer Accuracy |
± 3 mmHg or 2% of the reading whichever is greater |
Cycle Time in Auto mode |
1, 2, 3, 5, 10, 15, 30, 60, 90 Minutes (User selectable |
Cycle Time in Stat mode |
5 Minutes |
Thermal Recorder |
|
Paper Speed |
25 ± 1 mm/sec |
Paper Size And Type |
Thermal paper in rolls. 58 mm width, 15 meters |
Print Channel |
1 |
General |
|
Display |
7” Colour TFT, Resolution 800 x 480 |
Dimensions |
370 mm (W) x 300 mm (H) x 290 mm (D) |
Weight |
Less than 8.5 Kg. (including battery, external paddles) |
Operating Temperature Range |
0 to 40 ˚C at non-condensing RH between 15 and 95% |
Storage Temperature Range |
-10 to 60˚C at non-condensing RH between 20 and 80% |
AC Input |
100 – 240 Vac, 50/60 Hz |
Power Consumption |
< 240 VA |
Battery Type |
Sealed Lead Acid 12V, 4.5 AH, Rechargeable |
Battery Capacity |
> 100 discharges of 300 Joules |
Battery Charging Time |
Approx. 3 hours to 90% of full capacity. 18 hours to 100% capacity |
Event Information |
Last 24 critical events are stored along with ECG information. Events include MARK key press, Shock delivery and HR Alarm violation. |
Trend |
Last 200 NIBP measurement data and last 200 Alarm violation information |